MY-60
MY-60T
MY-61
MY-62
MY-63
MY-64
DIGITAL
MULTIMETER
OPERATOR’S
MANUAL
Seitenansicht 0
1 2 3 4 5 6 ... 16 17

Inhaltsverzeichnis

Seite 1 - OPERATOR’S

⊙ MY-60 ⊙MY-60T ⊙ MY-61 ⊙MY-62 ⊙ MY-63 ⊙MY-64 ○ ○ DIGITAL MULTIMETER OPERATOR’S MANUAL

Seite 2

1. SAFETY INFORMATIONS This multimeter has been designed according to IEC-1010 concerning electronic measuring instruments with an overvoltage catego

Seite 3

1.4 MAINTENANCE * Before opening the meter, always disconnect test leads from all sources of electric current. * For continue protection against fir

Seite 4

2.1 2.2 2.3 FUNCTION AND RANGE SELECTOR There are different functions and 32 ranges provided. A rotary switch is used to select functions as wel

Seite 5

The polarity of the red lead connection will be indicated along with the current value when making DC current measurement. 3. When only the figure”

Seite 6

3.6 TESTING DIODE 1. Connect the black test lead to COM jack and the red test lead to the VΩHz jack(The polarity of red lead is〝+〞) 2. Set the rotar

Seite 7

LOW BATTERY INDICATION SIZE (HxWxL) WEIGHT “”appears on the display 31.5mm×91mm×189mm 310g(including battery) 4.2 DC VOLTAGE Range Resolution Accura

Seite 8

4.6 RESISTANCE Range Resolution Accuracy 200Ω 2KΩ 20KΩ 200KΩ 2MΩ 20MΩ 200MΩ 0.1Ω 1Ω 10Ω 100Ω 1KΩ 10KΩ 100KΩ ±0.8% of rdg ± 3 digits ±0.8% of rdg ±

Seite 9

5.3 How to use the holster The holster is used to protect the meter and to make the measurement more comfortable. It comes with two stands installe

Seite 10

1. INTRODUCTION This MY-60 Series Digital Multimeter is a compact precision, battery operated, LCD display 3-½ digits Digital Instrument. Superiori

Seite 11

MY-60 MY-60T MY-61 MY-62 MY-63 MY-64200mV2V20V200V1000V0.8%±2AccuracyRange0.5%±10.5%±2 Input impedance: 10MΩ on all range 2) ACV MY-60 MY-60T MY-61 M

Seite 12

MY-60 MY-60T MY-61 MY-62 MY-63 MY-64200Ω2KΩ20KΩ200KΩ2MΩ20MΩ200MΩ1.0%±55.0%±20RangeAccuracy1.0%±101.0%±3 7) TEMPERATURE (MY-60T/MY-62/MY-64 only) MY-

Seite 13

measurement ranges. 6) Either one of the test leads should be taken off from the circuit under test when changing the test ranges. 7) To avoid elec

Seite 14

polarity at the red test lead connection will be indicated at the same time as the current. Note: a) If the current range is unknown beforehand, set

Seite 15

read the display value. Note: The tested capacitor should be discharged before the testing procedure. Never apply voltage to the “Cx” input socket, o

Seite 16

voltage approx. 2.8V. c) The meter displays the forward voltage drop and displays figure “1” for overload when the diode is reversed. 5.8 TRANSISTO

Seite 17

OPERATOR’S MANUAL CAUTION: * Before attempting to insert transistors capacitor thermocouple for testing, always be sure that test lest leads have bee